Color Calibration in Medical Imaging: Enhancing Consistency and Diagnostic Confidence

Abstract

This article examines the influence of color on medical imaging—even for images traditionally rendered in grayscale. With rapid advances in both imaging and display technologies, achieving consistent and correct color reproduction has become critical. Proper calibration ensures that diagnostic images are rendered faithfully, minimizes misinterpretation, and supports consistent performance across different devices.

Introduction

In modern medical imaging, accurate image reproduction is an essential requirement. Although many diagnostic images are shown in grayscale, even slight color shifts—such as a reddish or bluish tint—can affect how an image is perceived. The human eye does not perceive all shades uniformly, making consistent calibration vital for reliable diagnosis. In both clinical settings and remote consultations, ensuring that all devices display images with the correct tone and contrast is a top priority.

The Evolution of Color Imaging and Display Technologies

Increasing Prevalence of Color Images

Increasing Prevalence of Color Images

Influence on Medical Imaging

Ensuring Color Consistency

The primary goal of calibration is to achieve a consistent, device-independent color reproduction:

DICOM GSDF and L* Calibration Comparison

DICOM Calibration

CIE L* calibration

Balancing Grayscale and Color Requirements

The CSDF Hybrid Model and the Role of CIELAB

Many systems use what is sometimes referred to as the Color Softcopy Display Function (CSDF), a hybrid approach that combines DICOM’s grayscale calibration with standard color calibration methods. While this hybrid method provides a practical solution for devices that must display both grayscale and color images, it does not fully account for a significant advantage offered by the CIE Lab model.

The DICOM-based calibration is designed primarily to ensure consistent luminance levels using the GSDF. However, the CIELAB color space—particularly its L* component—offers a more perceptually uniform model of human brightness perception. In CIELAB, equal increments in L* are perceived uniformly by the human observer, an advantage that the CSDF does not fully capture. In essence, while CSDF is an effective blend of established protocols, adopting calibration methods based on CIELAB principles for the luminance channel could further enhance the perceptual accuracy and diagnostic reliability of medical displays.

Standards, Limitations, and Emerging Guidelines

Limitations in Existing Standards

Updated Guidelines and Recommendations

Calibration Strategies for Different Modalities

Grayscale Imaging on Color Displays

True Color Imaging on Color Displays

Practical Challenges and Future Directions

Practical Challenges and Future Directions

As display technologies continue to advance, the pursuit of accurate color and grayscale reproduction in medical imaging becomes even more vital. Robust calibration protocols—especially those that evolve from hybrid approaches such as the CSDF toward more perceptually accurate, CIELAB-based methods—are key to ensuring that clinicians receive diagnostically reliable images. This not only enhances diagnostic confidence but also supports improved patient outcomes in an increasingly digital healthcare landscape.

References:

AAPM TG18 report http://deckard.mc.duke.edu/ ~ samei/tg18

Digital Imaging and Communications in Medicine (DICOM)

Part 14: Grayscale Standard Display Function http://medical.nema.org/dicom/2004/04_14PU.PDF

Supplement 100: Color Softcopy Presentation Stateftp://medical.nema.org/medical/dicom/final/sup100_ft.pdf

CIE Space http://www.fho-emden.de/ ~ hoffmann/ciexyz29082000.pdf

ICC profiles http://color.org/

Ensure your mammography displays meet federal standards with our guide on MQSA and PerfectLum. Here, we break down what MQSA means, why it matters for your displays, and how to use PerfectLum to comply with these standards.

What Is MQSA?

The Mammography Quality Standards Act (MQSA) is a federal law enacted in 1992. Its purpose is to guarantee that every woman in the United States has access to high-quality mammography services for early breast cancer detection. To achieve this, MQSA:

Key requirements for facilities include:

For accreditation, facilities often work with trusted organizations like the American College of Radiology (ACR).

ACR Contact Information: The American College of Radiology 1891 Preston White Dr. Reston, VA 20191 1-800-227-6440

What Does MQSA Mean for Your Displays?

Under MQSA guidelines, mammography displays are vital components in the accurate evaluation of mammograms. These standards mean that:

Comprehensive Documentation: Detailed records of test data and any corrective measures are necessary for annual MQSA inspections.

MQSA Using PerfectLum

PerfectLum is designed to simplify your display QA process and streamline compliance. Here’s how:

1. Performing the QA Test

Use the integrated ACR Quality Assurance test available in PerfectLum:

  1. Setup:
    • Connect a colorimeter (such as the PerfectLum S1) to your display.
  2. Launch PerfectLum:
    • Open the PerfectLum application.
  3. Access QA Section:
    • Navigate to the Tpa Quality Assurance section.
  4. Select Testing Parameters:
    • Choose ACR as the testing method and Mammography as the category.
  5. Execute the Test:
    • Click “Perform Acceptance Test” and follow the guided wizard.

2. Data Synchronization and Reporting

PerfectLum is fully integrated with the PerfectLum QA Server. When connected, all QA test data are automatically synchronized to the server, ensuring a centralized, secure repository for all your test results. This seamless integration makes it easy to:

3. Preparing for the Annual MQSA Inspection

In addition to performing regular QA tests and synchronizing data, ensure you compile the following information:

4. Why Choose the ACR Test?

Using the ACR Digital Mammography Quality Control Manual for Full-Field Digital Mammography Systems (and its supplement for Digital Breast Tomosynthesis Mammography Systems) offers a proven alternative. Approved by the FDA as Alternative Standard #2423, this method provides:

Using PerfectLum’s integrated ACR testing and QA server synchronization, you can ensure that your displays continuously meet ACR requirements and streamline your compliance documentation.

An AAPM TG-270 display acceptance test establishes a baseline for a medical display before routine use. Identify its intended role, test it in the actual viewing environment, record the measurements and visual checks, then compare them with the facility’s approved criteria. AAPM Report 270 is the primary source for display QA recommendations.

PerfectLum can help run and document checks, but the software result is not a blanket declaration of clinical suitability. The medical physicist or display QA lead should select the appropriate category and review failures. For the grayscale target itself, see DICOM Part 14’s scope; for the product workflow, see PerfectLum.

Guidelines for Performing an Acceptance Test for AAPM TG 270 Using Qubyx PerfectLum

1. Introduction

AAPM Report 270 updates display QA guidance from the earlier AAPM TG-18 report. With the shift from CRT displays to modern flat-panel technology, TG 270 incorporates updated definitions of display types, revised test patterns, and performance standards that are tailored to contemporary medical imaging environments. It provides a framework for checking flat-panel displays used in medical imaging.

2. Understanding AAPM TG 270

What is AAPM TG 270?

Revised Performance Standards: Ensuring that displays used for image acquisition and review adhere to rigorous quality benchmarks.

3. Display Categories

AAPM TG 270 classifies displays into four distinct categories, each with its own performance criteria:

Electronic Health Record (EHR) Displays: Optimized for reviewing patient records and ancillary data, these displays balance clarity with ease of use.

4. How to Perform the AAPM TG 270 Acceptance Test Using Qubyx PerfectLum

The simplest and most efficient way to perform an acceptance test is by using Qubyx PerfectLum. Follow the steps below:

4.1. Preparation

4.2. Initiating the Test

  1. Launch the Application: Open Qubyx PerfectLum on your system.
  2. Access the Quality Assurance Tab: Navigate to the “Quality Assurance” tab within the application interface.
  3. Select the Regulation and Display Category:
    • Choose the AAPM TG 270 regulation.
    • Select the appropriate display category (Diagnostic, Modality, Clinical Specialist, or EHR).
  4. Enter Workstation Details: Fill in the required information regarding the workstation location.
  5. Connect Your Measurement Device: Attach your measurement device, such as the Qubyx X1, to ensure precise readings during the test.

4.3. Executing the Acceptance Test

  1. Start the Test: Click the “Perform Acceptance Test” button.
  2. Provide Tester Information:
    • Enter the tester’s name.
    • Record the ambient light conditions, as these can affect measurement outcomes.
  3. Follow the Wizard: The system will guide you through a series of visual tests. Answer all questions prompted by the wizard to complete the test process.

4.4. Review and Reporting

By following these comprehensive guidelines, you can easily perform an AAPM TG 270 acceptance test using Qubyx PerfectLum. The recorded results help the facility assess the display against its chosen criteria and document any corrective action.

For any technical questions or further assistance, please refer to the official TG 270 documentation or contact our support team.

Schedules
Diagnostic Mammography Diagnostic Modality Clinical Specialist

Electronic Health Record

Acceptance test measurement
Quantitative Min/Max Luminance Yes Yes Yes Yes Yes
Quantitative Luminance Response

Yes Yes Yes Yes Yes
Quantitative Color Assessment

Yes Yes Yes Yes Yes
Quantitative Ambient Luminance / Illuminance

Yes Yes Yes Yes Yes
Quantitative Uniformity

Yes Yes Yes Yes Yes
Acceptance Test visual
Qualitative Luminance response

Yes Yes Yes Yes Yes
Qualitative Ambient Luminance/Illuminance

Yes Yes Yes Yes Yes
Qualitative Spatial Resolution Yes Yes Yes Yes Yes
Qualitative Uniformity Yes Yes Yes Yes Yes
Diagnostic Mammography Diagnostic Modality Clinical Specialist Electronic Health Record
Consistency test measurement

Quantitative Min/Max

Annually Annually Annually Annually Annually
Luminance
Quantitative Luminance Response Annually Annually Annually No No
Quantitative Color Assessment Annually Annually Annually No No
Quantitative Ambient Luminance / Illuminance Annually Annually No No No
Quantitative Uniformity No No No No No

Consistency Test visual

Consistency Test visual
Qualitative Luminance response Quarterly Quarterly Quarterly Annually Annually
Qualitative Ambient Luminance/Illuminance Quarterly Quarterly Annually Annually Annually
Qualitative Spatial Resolution Quarterly Quarterly Annually Annually Annually
Qualitative Uniformity Quarterly Quarterly Annually Annually Annually

Reviewing category-specific thresholds

The tables below summarize suggested levels and checks from the 2019 AAPM TG-270 report (Tables VI and VII). Criteria depend on the display category and viewing conditions; a qualified reviewer must evaluate the combined measurements and select the facility’s acceptance plan. Verify these values against the primary report before clinical use.

Thresholds
Diagnostic Mammography Diagnostic Modality Clinical Specialist Electronic Health Record
Test measurement
Quantitative Min/Max Luminance LR target 350 (acceptable 250–450) L’min ≥ 1.2 cd/m2 L’max target 420 cd/m2 (acceptable ≥ 350 cd/m2). Note: L’max = Lmax + Lamb. L’min = Lmin + Lamb. LR= L’max/L’min LR target 350 (acceptable 250–450) L’min ≥ 1.0 cd/m2 L’max target 350 cd/m2 (acceptable ≥ 300 cd/m2). LR target 350 (acceptable 250–450) L’min ≥ 0.8 cd/m2 L’max target 250 cd/m2 (acceptable ≥ 200 cd/m2). LR target 350 (acceptable 250–450) L’min ≥ 0.8 cd/m2 L’max target 250 cd/m2 (acceptable ≥ 200 cd/m2). LR target 350 (acceptable 250–450) L’min ≥ 0.8 cd/m2 L’max target 250 cd/m2 (acceptable ≥ 200 cd/m2).
Quantitative Luminance Response DICOM GSDF deviation 10% DICOM GSDF deviation 10%

DICOM GSDF deviation 20% DICOM GSDF deviation 20% DICOM GSDF deviation 20%
Quantitative Color Assessment ΔD65 (u’ , v’) < 0.01 ΔD65 (u’ , v’) < 0.01 ΔD65 (u’ , v’) < 0.02 ΔD65 (u’ , v’) < 0.02 ΔD65 (u’ , v’) < 0.02

Quantitative Ambient Luminance / Illuminance

AR = Lamb/Lmin < 1/4 AR = Lamb/Lmin < 1/4 AR = Lamb/Lmin < 1/4 AR = Lamb/Lmin < 1/4 AR = Lamb/Lmin < 1/4
Quantitative Uniformity LUDM < 30% LUDM < 30% LUDM < 30% LUDM < 30% LUDM < 30%
Test visual
Qualitative Luminance response “Are all low contrast features, within ±5% contrast, visible under typical ambient conditions?” idem idem idem idem
Qualitative Ambient Luminance/Illuminance “Are the low-contrast features visible in the darkest areas of the test image under normal ambient light and with nearly dark conditions?” idem idem idem idem
Qualitative Spatial Resolution -“Does the graphics card to display pixel mapping correspond one-to-one (is the graphics card resolution identical to the native resolution of the display)?” idem idem idem idem
Qualitative Uniformity “There are no visible non-uniformities that would impact the clinical use of the display.” idem idem idem idem

Sicherung der Bildqualität in röntgendiagnostischen Betrieben – Teil 157: Abnahme- und Konstanzprüfung nach RöV an Bildwiedergabesystemen in ihrer Umgebung

Image quality assurance in diagnostic X-ray departments – Part 157: X-ray ordinance acceptance and constancy test of image display systems in their environment

Assurance de la qualité des images en diagnostic radiographique – Partie 157: Test d´acceptance et contrôle de constance selon RöV des systèmes de visualisation des images dans leur environnemen

DIN 6868-157

November 2014 – WHAT, WHEN and HOW it must be done

Conditions before testing (acceptance and constancy testing):

– The Image Reproduction System (IRS) has to be operated in the medical application, in which it is used later.

– Any modifications to the Image System must be recognizable. Those are for example changes to the application software, the display, the graphics card.

– The displays must be cleaned before starting the test.

– There should be no disturbing reflections on the screenfrom overhead lighting, windows, clothes.

– The room lighting must be stable during the test and when working.

– Before the test, the display must be warmed up according to the manufacturer’s instructions or at least 30 minutes.

– The DICOM curve must be selected or calibrated on all displays with diagnostic quality.

– The test images must match the resolution of the display.

– In case a quality assurance software is used for testing, it must be checked, whether the test images in the application software and in the quality assurance software are equally represented.

– The test shall be performed under the ambient conditions of the room class, which were determined by the radiation protection supervisor.

Test equipment (meters and test images)

Luminance Meters

For acceptance testing a calibrated telescopic luminance meter must be used for distance measurement.

To check consistency, a telescopic luminance meter (distance measurement), a photometer (contact-measurement Puck), a front integrated Meter (front sensor) or a backside integrated instrument can be used.

To measure the illuminance a special illuminance meter must be used.

Test Patterns

The following test patterns must be used and produced in the resolution of the display. A PIXEL of the testpattern must represent a PIXEL on the image reproduction device.  AAPM ->

TG 18-OIQ

TG 18-UN80

TG 18-MP

TG 18-LN8-01 to TG 18-LN8-18

TG 18-UN10

TG 18-LPH10

TG 18-LPH50

TG 18-LPH89

TG 18-LPV10

TG 18-LPV50

TG 18-LPV89

Room classes

Room classes are classified depending on use of the IRS. Depending on the room class different limits, testing and inspection intervals may be applied.

In general they are divided into two classes:

1. Image reproduction systems with diagnostic quality

2. Image reproduction systems with re-viewing quality

Furthermore, room classes are classified according to their requirements. The following table summarizes the room classes:

Room- class Room Activity Cassification Ambient illuminance
RC1 The radiology viewing room

Diagnosis of radiography and cross-sectional images

Diagnostic quality

≤ 50 Lx

RC2 Examination rooms with immediate diagnostic

An activity carried out in an examination room, where therapy-related decisions are made. Therefore, the image quality of the IRS must be sufficient for a diagnosis.

Diagnostic quality

≤ 100 Lx

RC3 Rooms for performing examinations

An examination is made with the help of a dialogue monitor. It is typically used for high contrast objects.

Diagnostic quality

≤ 500 Lx

RC4 Viewing and treatment rooms

Activities in which a known and evaluated medical diagnosis must be repeatedly retraced (e.g.

repositioning in the operating room, screw removal

Reviewing quality

≤ 1000 Lx

RC5 Dental diagnostic workplace

Diagnostics outside the lighting conditions of a dental treatment workplace.

Diagnostic quality

≤ 100 Lx

RC6 Dental Treatment

Room

Diagnostic under lighting conditions

of a dental treatment workplace.

Diagnostic quality

≤ 1000 Lx

The ILLUMINANCE is measured with the switched off image reproduction device and the measurement device

placed perpendicular to the screen surface

Acceptance test for image reproduction systems with diagnostic quality

The following two tables contain the necessary steps of the acceptance test.

Visual inspections

Testequipment (Test pattern) Test Partial test
TG18-OIQ

Overall image quality

unaltered visibility of line pair grid without streaks, smudges or erasures and visibility of the two PIXEL wide line pair grid with low CONTRAST in the center and in the four corners of the test pattern (test image

element 1)

RC5 (≤ 100 lx) and RC6 (≤ 1 000 lx): Visibility of 5 %- and 95 %- fields (test image element 2);

Visibility of the letters with low CONTRAST (”QUALITY CONTROL”

test image element 3);

   1) in the white and gray box: “QUALITY CONTROL” for all room classes;

   2) in the black field:

i)  Mammography (RC 1): „QUALITY CONTROL“

ii)  RC1 (≤ 50 lx): „QUALITY CONTRO“ iii)  RC2 (≤ 100 lx): „QUALITY CONT“ iv)  RC3 (≤ 500 lx): „QUALITY CON“

To calculate the maximum allowed defective pixels, multiply the horizontal number of PIXELs of image reproduction device with the vertical number of PIXELs and divide by 1,048,576 (equivalent to 1024×1024 pixels)

This result should be multiplied by the value specified in the table of

limit values and rounded off to the nearest whole number.

CLINICAL REFFERENCE IMAGES

Test based on CLINICAL reference images.

Not applicable

for RC5 und

RC6

For this test CLINICAL REFERENCE IMAGES are to be selected that have sufficient quality and image content in order to evaluate correctly the image quality of the display.

TG18-LPH10,

-LPH50 und

-LPH89 (horizontal Lines);

TG18-LPV10,

-LPV50 und

-LPV89 (vertikal Lines.)

The visibility of the line structures in the test pattern 6. Required only

for RC1 and

Mammography

The visibility of the line structures in the test pattern 6 must be checked for RC 1.

TG18-LPH10, TG18-LPH50 und TG18-LPH89 (horizontal Lines) TG18-LPV10, TG18-LPV50 und TG18-LPV89 (vertikal Lines)

Metrological (measurement) tests for acceptance testing

Testequipment (Test pattern) Test image and Method of measurement Test Test Description
Luminance

Meter (Distance measurement)

TG18-LN8-01

Minimum

Luminance

The minimum luminance L min is measured on the screen center with a telescopic measuring device.

TG18-LN8-18

Maximum

Luminance

The maximum luminance L’max is measured on the screen center with a telescopic measuring device.

Luminance measurement. Measured with

the switched off display

Veiling

Luminance

The veiling luminance Lamb is measured on the screen center with a telescopic measuring device. The display must be switched off.

Luminance Meter (Contact measurement)

TG18-UN80 and TG18-UN10

Homogeneity

of the display

It must be measured (for 8 bit corresponding DDL DDL 26 and 204) in essay measurement at five points or a screen diagonal of ≥ 23 inches at nine points in the luminance of 10% and 80% of the maximum DDL. The UNIFORMITY H is calculated as the percentage difference between the highest (Lhighest) and lowest (Llowest) values of the

measured LUMINANCE with respect to its mean as follows:

H = 200 ×(Lhighest – Llowest)/(Lhighest + Llowest)

TG18-UN10

Homogeneity of multiple

displays

It is measured in the screen center (z) of all the displays connected to the same workstation in order to determine the uniformity between the displays. The UNIFORMITY of multiple display devices is calculated as the percentage difference between the highest and lowest values of the luminance in relation to their mean value:

H = 200 × (Lhighest – Llowest)/(Lhighest +

Llowest)

TG18-LN8-01 to

TG18-LN8-18

Luminance Responce Curve – DICOM GSDF

The LUMINANCE RESPONSE CURVE is determined by representing the 18 test pattern TG18-LN8-01 to

TG1-LN8-18 and measurement on the screen center. The evaluation of the measurements must be made with a suitable software tool. The veiling luminance must be considered while calculating.

Illuminance meter

With a switched off display, perpendicular to the screen surface

(screen center)

Illuminance The illuminance is measured while the image reproduction device is switched off and the measurement device placed perpendicular to the screen surface.

Minimum requirements for image display devices (excluding mammography)

The body region/ methods Max Display Luminan- ce cd/m2 Max Lumi- nance ratio r’= L’max/ L’min Matrix/ resolu- tion of the display Lumi- nance respon- ce curve deviation from the DICOM GSDF curve Homo- genity within the display Homo- genity bet- ween several displays Pixel- errors (per 1024x 1024) depen- ding on the error type) Minimal Luminance cd/m2 Minimal pixel size
Projection radiology (Thorax,

skeleton, abdomen)

≥ 250

≥ 250

≥ 1600

x

≥ 1200

GSDF

± 10%

<25%

<20%

A ≤ 1

B ≤ 5

C ≤ 5

D≤ 1

≤ 1.1 x Lam 140 micro meter
Fluoroscopy,

all applica- tions

≥ 150 ≥ 100 ≥ 1024 x

≥ 1024

GSDF

± 15%

<25% <20%
Computed tomog- raphy(CT) ≥ 150 ≥ 100

≥ 1024 x

≥ 1024

GSDF

± 15%

<25% <20%

For RC 5:

Dental

digital volume tomography, intraoral

x-ray diagnos

tic with dental x- ray tube heads, panoramic radiographs, cephalometric radiographs of

the skull, dental skull

plain films, hand- held shooting for skeletal growth determination

≥ 200 ≥ 100 ≥ 1024 x

≥ 768

only visual test

with the

test image TG18-O IQ and the

test

image

elements e a), b), c), e)

and g)

<30% <30%

For RC 6:

Intraoral

X-ray diagnos

tics

with dental

x- ray tube heads, panoramic radiographs, cephalometric radiogr phs of the skull, dental skull plain films, hand- held shooting for skeletal growth determination

≥ 300 ≥ 100 ≥ 1024 x

≥ 768

only visual tests with

the test image

TG18-O IQ and

the test

image elements e a), b),

c), e)

and g)

<30% <30%

Minimum requirements for image display devices for the mammography. In addition the following test patterns must be

checked for the visibility of the line structures: TG18-LPH10, -LPH50 und -LPH89 (horizontal Lines) TG18-LPV10, -LPV50 und -LPV89 (vertical Lines)

The body region/ methods Max Display Luminan- ce cd/m2 Max Lumi- nance ratio r’= L’max/ L’min Matrix/ resolu- tion of the display Lumi- nance respon- ce curve deviation from the DICOM GSDF curve Homo- genity within the display Homo- genity bet- ween several displays Pixel- errors (per 1024x 1024) depen- ding on the error type) Minimal Luminance cd/m2 Minimal Luminance cd/m2
Mammo

graphy

≥ 250

≥ 250

≥ 2048 x

≥ 2048

GSDF

± 10%

<25%

<10%

A ≤ 1

B ≤ 5

C ≤ 5

D≤ 1

≤ 1.1 x Lam 140 micro meter
Mammo

graphic

Stereota-

xy

≥ 200

≥ 100

≥ 1024 x

≥ 1024

GSDF

± 15%

<25% <20%

Reference values for constancy tests

If the acceptance test passed, the measured values must be stored as reference values for constancy tests.

In case of a deviation of the measured values by more than 20% from the

reference values of the ACCEPTANCE TEST it is recommended to take an appropriate action in order to prevent further deviations. In case of deviations of the measured values by more than 30%, it is required to correct the errors.

After the error has been removed, a CONSTANCY or an ACCEPTANCE TEST is

required. If the significant changes had been made to the image reproduction device during troubleshooting, an acceptance testing is mandatory.

Constancy tests at workplaces with diagnostic quality

The constancy tests must be carried out at the intervals indicated below or after significant changes in the workstation.

If the veiling luminance is less than 0.15 cd /m2, a negative deviation from the reference value must not be considered.

Visual steps for constancy tests Interval

Test equipment (Test Image)

Test

Test Description

TG18-OIQ Overall image quality Unaltered visibility of line pair grid without streaks, smudges or cancellations and visibility of the two pixel wide line pair grid with low contrast in the center and in the four corners of the TEST IMAGE (test pattern element 1)

Daily
RC5 (≤ 100 lx) and RC6 (≤ 1000 lx): In the test pattern element 2 (5% – and 95% field) the low-contrast squares must be recognizable.

Visibility of the letters with low CONTRAST (“QUALITY CONTROL” test image element 3);

1)In the white and gray box: “QUALITY CONTROL” for all room classes

2)  in the black field:

i)  Mammography (RC1): „QUALITY CONTROL“

ii)  RC1 (≤ 50 lx): „QUALITY CONTRO“

iii)  RC2 (≤ 100 lx): „QUALITY CONT“

iv)  RC3 (≤ 500 lx): „QUALITY CON“

The following applies for room class 5 and 6: At the test pattern element 3 at least in the mid gray box the words ”QUALITY CONTROL” must be visible.

Continuity of the appearance of the progress bar (test image element

7);

TG18-UN80

Luminance uniformity

Visual assessment of the test pattern for interfering irregularities from the center to the edges.

Semi-

annually

Color appearan-

ce and uniformity

The color uniformity over the screen of the displayed test pattern is to be verified. The test must be performed on multiple image display devices of the same type that are connected to the same image display system and are installed next to each other. This is to test for perceptual color differences between the individual IMAGE DISPLAY DEVICES.

Metrological (measurement) tests for constancy tests. The same limits as for the acceptance test must be applied

Interval: Semi-annual

Test Equipment Test image and method of measurement Test Test Description
Luminance Meter (Contact measurement and distance measurement)

TG18-LN8-01

Minimum

Luminance

The minimum luminance L min is measured on the screen center with a telescopic measuring device.

TG18-LN8-18

Maximum

Luminance

The maximum luminance L’max is measured on the screen center with a telescopic measuring device.

Luminance measurement. Measured with

the switched off display

Veiling

Luminance

The veiling luminance Lamb is measured on the screen center with a telescopic luminace meter with the display turned off. The test is only required if there were differences in the minimum luminance.

The test is not required for room classes 3,

5 and 6 .

Luminance Meter

(Contact

measurement and distance measurement)

TG18-UN10

Homogenity of multiple

displays

Is is to be measured in the screen center (z) of all the displays connected to the same workstation, in order to determine the uniformity between the displays.

The uniformity of multiple displays is calculated as

the percentage difference between the highest and lowest values of the measured LUMINANCE taking into account its mean as follows:

H = 200 × (Lhighest – Llowest)/(Lhighest +

Llowest)

The test is not required for RC3, RC5 und RC6.

Luminance Meter

(Contact

measurement)

TG18-LN8-01 to

TG18-LN8-18

Luminance curve DICOM GSDF

The LUMINANCE CURVE is determined by representing the 18 TEST PATTERNs  TG18 from TG18 to-LN8-01 to LN8-18. Measurements must be taken on the screen center using a contact measurement method.

The evaluation of the measurements must be made with a sutable software tool.

The veiling luminance must be considered while calculating.

The test is not required for RC3, RC5 und RC6.

Illuminance meter

with a switched off display measurement device is placed perpendicular to the screen surface (screen center) Illuminance The illuminance is measured while the image reproduction device is switched off and the measurement device placed perpendicular to the screen surface.

Terms and definitions

Acceptance test

Examination of an image display system (workstation), including the image reproduction device (display) immediately after installation and before operating assumption. Measured values are stored as reference values    for constancy tests.

Application software

The software that is used to display the medical images.

Artifact

The structures that can be recognized by the viewer but are not part of the image.

Diagnostic quality

Visual quality of the image display systems that are used to allow diagnostic statements and used as the basis of medical decisions

Illuminance

E

The strength of the luminous flux incident on a surface per unit area. Unit: Lux (lx).

Viewing quality

Visual quality of the image reproduction systems that are used in order to view the already diagnosed images to detect  image features and content.

IMAGE REPRODUCTION DISPLAY (in the DIN named as “BWG”)

In general, a monitor. Or other displays as projectors.

Image Reproduction System (in the DIN named as “BWS”)

A workstation consisting of the computer graphic card, application software and all connected displays.

Grayscale Standard Display Function GSDF

Function for reproduction of grayscales on a computer display. Generally known as DICOM curve, DICOM GSDF curve Part 14 or Barten curve.

Luminance Uniformity

H

Uniformity of brightness distribution on one or multiple displays.

Clinical reference image

A medical image that is typical for the application of image reproduction system, and contains important image elements that allow conclusions about the quality of the image.

Constancy test

Checks carried out at regular intervals in order to identify and document changes compared to the reference values    of the acceptance test.

Luminance

L

The luminance describes the brightness of extended, planar light sources. Unit: cd/m2  (candela per m2)

Luminance coefficient

q

Quotient of the LUMINANCE of the surface element in the given direction and ILLUMINANCE

on the medium.

Maximum display luminance

L’max

Maximum luminance of the display.

Maximum Luminance Ratio

r’

Ratio of MAXIMUM Luminance L’max and MINIMAL LUMINANCE L’min.

r=’ L’max/ L’min

Maximum Luminance

L’max

Maximum luminance in the center of the screen plus veiling luminance L’max= Lmax+ Lamb

Minimal Display Luminance

Lmin

Minimal luminance of the display.

Minimal Luminance

L’max

Minimum luminance in the center of the screen plus veiling luminance.

L’max  = Lmax + Lamb

Veiling Luminance

Lamb

Luminance generated by the ambient light on the surface of a display (reflection of ambient light on the display)

The ACR–AAPM–SIIM technical standard discusses how display performance and quality control fit into medical imaging. For a practical review, start with the display’s intended clinical use, the room lighting, luminance response, spatial performance, and a documented QA plan. The current ACR–AAPM–SIIM standard is the primary reference.

What should an ACR display QA review include?

The detailed tables, screenshots and operating-system lists below reflect an earlier article summary and may not match the current edition or product release. Confirm numerical criteria against the current standard and your facility’s medical physicist; check the downloads page for current installers. For a complementary test workflow, see our AAPM TG-270 acceptance-test guide; PerfectLum can help document display measurements and QA.

ACR–AAPM–SIIM TECHNICAL STANDARD FOR ELECTRONIC PRACTICE OF MEDICAL IMAGING

Part I

This standard is a set of guidelines created to assist practitioners in providing appropriate radiologic care for patients.

It applies to any system of digital image data management except digital mammography.

The standard is divided into the following sections:

I. INTRODUCTION

II. QUALIFICATIONS AND RESPONSIBILITIES OF PERSONNEL,

III. EQUIPMENT SPECIFICATIONS,

IV. DOCUMENTATION,

V. LICENSING, CREDENTIALING, AND LIABILITY,

VI. RADIATION SAFETY IN IMAGING,

VII. QUALITY CONTROL AND IMPROVEMENT, SAFETY, INFECTION CONTROL, AND PATIENT EDUCATION

Our focus will be the information pertaining to display performance and its assessment, as well as its quality control guidelines  (Sections III and VII), but it is important to analyze it within the entire diagnostic imaging paradigm including acquisition, compression, transmission, archiving, retention etc to get a full picture.  

III. Equipment Specifications

 The equipment used in digital image data management should provide image quality and availability appropriate to the clinical needs (official interpretation or secondary review).

According to the standard, for newly acquired equipment compliance with the following standards is recommended:

Subsections A – C of the ACR describe image acquisition, compression and transmission, while subsection D focuses on its display, so we are going to look into it more closely.

First of all, images seen by all hospital staff providing patient care (technologists, radiologists, physicians) should have similar appearance. This means that each monitor involved in this chain should display the image in the same way, which can be achieved by calibrating it with a special software.

Since the way  images are displayed  is influenced by numerous factors, the ACR outlines the requirements to the workstation and display performance, which can be summarized as follows:

1. Workstation characteristics

Parameter Recommendations
Graphic bit depth 8 bits (256 values)

Display technology Before purchase the viewing angle characteristics of the LCD device should be evaluated using contrast transfer test patterns.

Graphic interface

DVI-D (either single-link or dual-link) or display port.

Image presentation size

the interpolation of each displayed pixel, whether up-sampling or down-sampling, should consider more than the closest four acquired pixel values

Presentation support features

The software used to select and present images should allow quick and easy review or interpretation of a study and include such features as:

  • fast and easy navigation between new and old studies
  • window and level adjustment tools
  • zoom (magnification) and pan function etc.

2. Display characteristics: recommendations and thresholds

a. Luminance response

Parameter Explanation Recommendations / Thresholds
Ambient luminance (Lamb) the brightness shown by the display surface when the power to the display device is off

should be less than one-fourth of the luminance of the darkest gray level.

Minimum luminance (Lmin) the luminance of the lowest gray value including a component from ambient lighting, L’min = Lmin + Lamb

for diagnostic interpretation – at least 1.0 cd/m2

for the interpretation of mammograms – 1.2 cd/m2

for other uses – 0.8 cd/m2

Maximum luminance (Lmax) the luminance for the maximum gray value

for diagnostic interpretation – at least 350 cd/m2

for the interpretation of mammograms – at least 420 cd/m2

for other uses – at least 250 cd/m2

Luminance ratio

(LR)

the ratio of Lmax to Lmin all display devices in a facility should have the same LR (greater than 250)
White point

can have different shades; the standard recommends D65, which corresponds to daylight, indirect sunlight at noon

monitors should be set to a white point corresponding to the CIE daylight standard D65 white point. This corresponds to a color temperature of about 6,500 degrees K

b. Pixel pitch and display size: recommendations and thresholds

Parameter Explanation Recommendations / Thresholds
pixel pitch

the spacing of pixel structures

for diagnostic interpretation – not larger than 0.210 mm ( 0.200 mm recommended)

for monitors used by technologists and clinical care staff – not larger than 0.300 mm (0.250 mm recommended)

display size

diagonal the diagonal display distance should be about 80 percent of the viewing distance

At 2/3 meter, this corresponds to a diagonal size of 53 cm (21 inches)

aspect ratio

width to height of the display

3:4 or 4:5

16:9 or 16:10 (can be used if the application software can present images in two regions with 8:9 or 8:10 aspect ratio)

Section III E-H of the ACR provides information concerning Archiving, Retention, and Retrieval; Image Sharing; Security, Privacy, Reliability, and Redundancy; Informatics Infrastructures and Workflow Processes;

Section IV outlines the recommendations for making reports by physicians  officially interpreting examinations  using digital image data management systems, and

Sections V and VI are about licensing, credentialing, and liability (V) and radiation safety in imaging (VI).

We will look into Section VII (Quality control and improvement, safety, infection control, and patient education) more closely, namely into the part dealing with quality control.

Quality Control

The recommended tests can be subdivided into Visual and Measurement tests.

Visual Tests (aimed at evaluating the quality of image display visually, without taking measurements):

Parameter Explanation Recommendations / Thresholds
1. the overall operation of the system under conditions that simulate its normal operation (for acquisition workstation and secondary display devices)

a test image (such as the AAPM TG18-QC test pattern) should be captured, transmitted, archived, retrieved, and displayed

at appropriate intervals

2. a spatial resolution test

at least 2.5 lp/mm resolutions for official interpretation

display monitors used for primary interpretation: monthly

3. a display fidelity test

TG18-QC pattern data files sized to occupy the full area used to display images on the monitor should be displayed. The overall SMPTE image appearance should be inspected to assure the absence of gross artifacts (e.g., blurring or bleeding of bright display areas into dark areas or aliasing of spatial resolution patterns). As a dynamic range test, both the 5% and the 95% areas should be seen as distinct from the respective adjacent 0% and 100% areas.

display monitors used for primary interpretation: monthly

4. contrast response

can be done using a visual test pattern for evaluating contrast response

display monitors used for primary interpretation: monthly

5. the viewbox luminance

should be sufficient to meet the diagnostic needs of the imaging procedure

applicable industry standards and/or recommendations should be followed when available.

Measurement test (needs to be done with the special measurement devices):

Parameter Procedure Frequency
6. luminance response – DICOM curve

measure the luminance in relation to gray value and evaluate the contrast

quarterly

Part II

Routine quality assurance (QA) is necessary and essential to ensure the performance of medical displays. As mentioned above, according to the ACR guidelines, the evaluation includes visual (spatial resolution, display fidelity, contrast response, the viewbox luminance) and measurement (luminance response to the DICOM GSDF) tests. However, this manual process can be very time consuming. The second part of this paper describes how a tool can support the ACR-related QA workflow and document measurements, which aims to increase physicist efficiency, and improve the consistency of QA results by reducing human error.

The following seven steps describe a PerfectLum workflow; the facility must review the measurements against its applicable acceptance criteria.

PerfectLum Client Technical Specifications and Compatibility

Windows Mac OS X
  • XP (32-bit and 64-bit)
  • Vista (32-bit and 64-bit)
  • Win 7 (32-bit and 64-bit)
  • Win 8 (32-bit and 64-bit)
  • 10.6 Snow Leopard (32-bit and 64-bit)
  • 10.7 Lion
  • 10.8 Mountain Lion
  • 10.9 Mavericks
  • 10.10 Yosemite
  • Processor: Pentium or AMD K7
  • Minimum RAM: 512MB.
  • Minimum free hard disk space: 150 MB.
  • Intel hardware is required
  • Processor: Intel only. No
  • PowerPC support.
  • Minimum RAM: 512MB.
  • Minimum free hard disk space: 150 MB.

Display compatibility: All LCD, CRT displays and projectors.

Required equipment: measuring device (puck). Please refer to the list of supported measuring devices for detailed information.

Step 1. Download and install the software

The Qubyx PerfectLum software is available for download from https://qubyx.com/en/download

Have troubles with installation? Please refer to PerfectLum User Manual or contact our support team at support@qubyx.com 

Step 2. Set up PerfectLum in accordance with the guidelines from ACR

After you successfully installed the software, you will need to set it for the ACR regulation.

This can be easily done by selecting the corresponding quality assurance settings. Please go to PerfectLum Settings – > Preferences and do the following:

In the “Calibration Settings” menu enter the values required by the ACR guidelines:

In the “QA Settings” menu:

Figure 1: Quality Assurance Settings

In the “Display Settings” menu:

Step 3. Measure ambient light

Go to PerfectLum Suite Settings – > Preferences – > Calibration Settings and click the “Measure Ambient Light” button. During this measurement the display needs to be switched off, as we will measure the reflection of the ambient light on the display panel. This way the veiling glare is measured, which is the brightness reflected by the display surface. During the measurement process the software will switch off the backlight of the panel. You must place the sensor 15 cm away from the display facing the switched off panel.

IMPORTANT: During the measurement process don’t move the mouse and don’t hit any keys as it will switch the display back on. Once the measurement is finished the display will be automatically switched back on.

Now you are ready to perform the calibration.

IMPORTANT: the ambient light measurement must be taken in the same lighting conditions as during the normal display operation and the subsequent calibration and acceptance tests; otherwise the display can fail the acceptance test.

Step 4. Calibrate your display to DICOM GSDF

To start the calibration, please select the display you want to calibrate from the menu on the left and click the “Calibration” button.

Figure 2. PerfectLum main window

IMPORTANT: Before starting the calibration, please warm up your display for at least 30 minutes.

Place the measuring device on the display and start the calibration process. Make sure no ambient light reaches the sensor. After calibration is finished, PerfectLum will provide a conformance report.

Step 5. Perform the QA acceptance test

The ACR acceptance test consists of two parts: a set of visual tests and a measurement part (a sensor is needed). To start the “Acceptance test” click “Acceptance Test” in the PerfectLum main window and simply follow the instructions given by the test wizard.

IMPORTANT: If you are running a demo version of the product, you will not get the detailed acceptance test report for your display. The generated report is a sample and does not show the real results of the performed test.

Figure 3. Automated acceptance test

The ACR guidelines require to perform visual constancy tests monthly and measurement constancy tests quarterly.

After the acceptance test is performed, PerfectLum will automatically generate schedules for the corresponding constancy tests.

Figure 4. PerfectLum Scheduler

Step 6. Perform Constancy tests when they are due

PerfectLum will notify you when the tests are due. Click the “Perform Scheduled Tasks”’ button to carry out a scheduled task or “Remind Me In” to choose when you will be reminded again. When performing the tasks, simply follow the Test Wizard.

Figure 5. PerfectLum Test Wizard

Step 7. Get the ready-to-print pdf reports

PerfectLum generates a PDF report for each completed test that can be printed using a PDF reader. Each report will be saved in the history database.

Have questions?

Contact us at support@qubyx.com

Qubyx Software TechnologiesLTD

HK

qubyx.com

Glossary

1. Analog signal – a form of information transmission in which the signal varies in a continuous manner and is

not limited to discrete steps.

2. Archive – a repository for digital medical images in a picture archiving and communications system (PACS), typically with the specific purpose of providing either short-term or long-term (permanent) storage of images. Erasable or nonerasable media may be used in an archive.

3. Bit (binary digit) – the smallest unit of digital information that a computing device handles. It represents off or on (0 or 1). All data in computing devices are processed as bits or strings of bits.

4. Bit depth – the number of bits used to encode the signal intensity of each pixel of the image.

5. Compression ratio – the ratio of the number of bits in an original image to that in a compressed version of that image. For example, a compression ratio of 2:1 would correspond to a compressed image with one-half the number of bits of the original.

6. CR (computed radiography) – a system that uses a storage phosphor plate contained in a cassette instead of a film-screen cassette. A laser beam scans the exposed plate to produce the digital data that is then converted into an image.

7. CRT (cathode ray tube) – an older technology monitor or display device used for viewing digital softcopy images. A CRT uses a controlled beam of electrons incident on a phosphor to generate a luminous image.

8. Data communication – all forms of computer information exchange. Data communication may take place between two computers in the same building via a local area network (LAN), across the country via telephone, or elsewhere by a wide-area network (WAN).

9. Data compression – methods to reduce the data volume by encoding it in a more efficient manner, thus reducing the image processing and transmission times and storage space required. These methods may be reversible (lossless) or irreversible (lossy).

10. DICOM (Digital Imaging and Communications in Medicine) – a standard for interconnection of medical digital imaging devices, developed and sponsored by the American College of Radiology and the National Electrical Manufacturers Association, consisting of a standard image format and a standard communications protocol.

11. Digital signal – a form of information transmission in which the signal varies in discrete steps, not in a continuous manner.

12. Digitize – the process by which analog (continuous value) information is converted into digital (discrete value) information.

13. Direct image capture – the capture or acquisition of digital image data that have been acquired in digital format by an imaging modality. The image produced from the data, regardless of the modality that produced it (CT, MRI, CR, US), should include the full spatial resolution and bit depth of the original.

14. Dynamic range – the difference in signal intensity, or frequency, between the largest and smallest signals a system can process or display. Increasing the number of bits per pixel in a digital image increases the dynamic range of the image.

15. File – a set of digital data that have a common purpose, such as an image, a program, or a database.

16. Grayscale – the number of different shades of levels of gray that can be stored and displayed by a computer system. The number of gray levels is directly related to the number of bits used in each pixel: 6 bits = 64 gray levels, 7 bits = 128 gray levels, 8 bits = 256 gray levels, 10 bits = 1,024 gray levels, and 12 bits = 4,096 gray levels.

17. Hardware – a collective term used to describe the physical components that form a computer. The monitor, CPU, disk drives, memory, modem, and other components are all considered hardware.

18. Image matrix size – The size of an image described as the number of rows and the number of columns of pixels.

19. Image sampling: Up-sampling – Small matrix images are typically sampled more finely than the acquired pixel spacing in order to increase the number of rows and columns and increase the presented size. Down-18 / Electronic Practice TECHNICAL STANDARD sampling – Large matrix images are typically sampled more coarsely than the acquired pixel spacing in order to decrease the number of rows and columns so that the full image area can be presented.

20. Irreversible compression – some permanent alteration of digital image data. This is sometimes referred to as lossy compression.

21. LCD (liquid crystal display) – a modern monitor or display device used for viewing digital softcopy images.

22. Lossless compression – see reversible compression.

23. Lossy compression – see irreversible compression.

24. Monochrome monitor – a computer display in which an image is presented as different shades of gray from black to white (see also grayscale).

25. Mouse – an input device that allows the computer user to point to objects on the screen and execute commands.

26. PACS – Picture Archiving and Communication System.

27. Phosphor – the coating on the inside of a CRT or monitor that produces light when it is struck by an electron beam.

28. Pixel (picture element) – the smallest piece of information that can be displayed on a CRT. It is represented by a numerical code within the computer and displayed on the monitor as a dot of a specific color or intensity. An image is composed of a large array of pixels of differing intensities or colors.

29. Protocol – a set of guidelines by which two different computer devices communicate with each other.

30. RAM (random access memory) – a type of temporary memory in a computer in which programs are run, images are processed, and information is stored. The amount of RAM that a computer requires varies widely depending on the specific application. Information stored in RAM is lost when the power is shut off.

31. Resolution – spatial resolution is the ability to distinguish small objects at high contrast. It is related to and in some cases limited by the pixel size. Contrast (grayscale) resolution is the ability of a system to distinguish between objects of the same size having different signal intensity. It is related to and in some cases limited by the bit depth.

32. Reversible compression – no alteration of original image information upon reconstruction. This is sometimes referred to as lossless.

33. Secondary image capture – the capture in digital format of image data that originally existed in another primary format (e.g., a digital image data file on a CT scanner, or a screen-film radiographic film) through the process of video capture or film digitization.

34. SMPTE – the Society of Motion Picture and Television Engineers.

35. Software – a name given to the programs or sets of programs that are executed on a computer

PerfectLum Suite: effortless compliance

with the NY PDM guide

This detailed and guide will help you make sure your display complies with the NY PMD regulation while using PerfectLum Suite. Follow these simple steps:

STEP 1. Install PerfectLum Suite.

Pre-installation conditions:

Power on the workstation with all display(s) attached that you intend to use. Check if the screen settings are correctly configured:

XP – Go to Settings -> Control panel, and double click on the display icon. Select the “Settings” tab to view screen resolution and color settings.

Vista – Go to Settings -> Control panel and click on “Adjust screen resolution” under the Appearance and Personalisation Icon.

Windows 7 and 8 – Go to Control panel -> Display -> Adjust resolution.

Mac OS X – System Preferences -> Displays and choose resolution radio box “Best for display”.

Ensure that the screen resolution is set to the native resolution of the display with color quality of 32 bit “True color” or higher. Ensure that the correct graphics card display driver is installed. Repeat these checks for each display attached to your system. Your system is now ready to start the installation process.

IMPORTANT: Before starting the installation process, please make sure that the USB measurement device is NOT connected. The measurement device will be added later, after PerfectLum Suite is successfully installed.

IMPORTANT: If you have any other calibration software installed, please disable or uninstall it, as it can interfere with the proper operation of PerfectLum Suite.

PerfectLum Suite can be operated as a client only or as a client-server combination. Complete installation consists of several components:

  1. PerfectLum Suite Client is the client application that runs on each of the workstations. It provides a number of actions that can be performed locally on the workstations, including calibrating the displays, viewing test patterns or performing full QA (Quality Assurance) and acceptance tests (QA tests available only in PerfectLum Suite). Even when PerfectLum Suite Client is not connected to PerfectLum Remote Server it can still run all QA tasks on the workstation.
  1. PerfectLum Remote Server. All PerfectLum Suite clients can be connected to PerfectLum Remote Server. No special license is required. QA managers can connect to PerfectLum Remote Server via internet/intranet connection and take full control over the performance of connected workstations from any location at any given time.

Installation

Follow the instructions presented in the installer and click on the QUBYX software license agreement after reading it carefully.

If you don’t want to install PerfectLum Suite into the default directory, browse your computer and choose another one. Windows will present a driver message for a few times during PerfectLum Suite installation. This is normal behavior as drivers to support measurement devices are loaded. Please click the „Install this driver software anyway‟ button when prompted.

Proceed by double clicking on the “box”.

Follow the instructions presented in the installer and click on the QUBYX software license agreement after reading it carefully.

Click “Install” to begin the installation.

STEP 2. Install your Measurement Device

You will need a measurement device during the tests involving measurements (Luminance Response and Luminance Response – DICOM Curve). PerfectLum Suite is compatible with any display and the following measurement devices:

Display Front Sensor

Colorimeter

Spectral photometer

Spot Luminance meter

In order to install it:

Plug the device into an unused USB port and refer to the following depending on which OS you have installed the software:

Vista and Windows 7 Vista and Windows 7 will automatically install your device driver. Wait until you receive the “Device driver software installed successfully‟ message, and jump ahead to the end of this section.

Windows XP: The “Found new Hardware Wizard” will start. Click the “Yes, this time only” radio button followed by “Next”.

Check the “Install the Software Automatically (Recommended)” radio button and click “Next”.

The software is installed automatically.

When all drivers have been loaded and configured, the wizard will prompt you to click on the “Finish” button.

Windows will notify you when the measurement device is ready to use, by momentarily displaying the “Found New Hardware” message balloon in the lower right corner of the screen.  

STEP 3. Activate your license

To activate the Calibration and QA license, click the “License” option in the main window, and select “Registration/Activation”. When the Registration/Activation window opens, enter the serial number you obtained when you purchased a license (see your invoice), and choose an activation type.

For automatic online activation, just enter and repeat your e-mail address, and click the “Activate” button.

For manual activation, follow the link and fill out the form that you will find on the website. Enter the generated workstation ID that you see below into the corresponding field of the form. When you submit the form, an activation code will be sent to your e-mail. Enter this code into the field below and click “Activate”.

When you finish the license activation, all licenses in the Preferences > License and Licensing section will be marked as active.

STEP 4. Set up PerfectLum for NY PDM

This can be done by selecting the corresponding quality assurance settings.

Go to PerfectLum Suite Settings > Preferences and do the following:  

In the “Calibration Settings” menu enter the values in accordance with the NY PDM requirements:

In the “QA Settings” menu:

In the “Display Settings” menu:

STEP 5. Measure ambient light

Go to PerfectLum Suite Settings > Preferences > Calibration Settings and click the “Measure Ambient Light” button. During this measurement the display needs to be switched off, as we need to measure the reflection of the ambient light on the display panel. This way the veiling glare is measured, which is the brightness reflected by the display surface. During the measurement process the software will switch off the backlight of the panel. You must place the sensor 15 cm away from the display facing the switched off panel.

During the measurement process don’t move the mouse and don’t hit any keys as it will switch the display back on. Once the measurement is finished the display will be automatically switched back on.

Now you are ready to perform your first calibration.

IMPORTANT: the ambient light measurement must be taken in the same lighting conditions as  during the normal display operation and the subsequent calibration and acceptance tests; otherwise the display can fail the acceptance test.

STEP 6. Calibrate your display

First select the display you want to calibrate from the menu on the left. After that, click the “Calibration” button.

IMPORTANT: Calibration should always be started on a warmed-up display. Warm-up time is specified by the display manufacturer (minimum 30 minutes).

Place the sensor on the display and start the calibration process. Make sure no ambient light reaches the sensor. After calibration is finished a report will pop up on the screen with the calibration results.

STEP 7. Perform an Acceptance test

The acceptance test for NY PDM consists of two parts: a measurement part (DICOM GSDF verification test and Luminance Response test) and a set of visual tests (Visual verification test, White Uniformity test, Black Uniformity test, Color distortion test, All displays White Uniformity test, All displays Black Uniformity test).

For measurement tests an on-screen or a handheld measurement device must be used.

To start the “Acceptance test” click “Acceptance Test” in the PerfectLum main window.

You will be asked to fill in the tester’s name and the ambient conditions (here you’ll need to describe the ambient conditions of the workstation).

Answer the questions prompted by the Wizard  to evaluate the display performance visually.

When the measurement part starts, follow the on-screen instructions.

Please note that during the ambient light measurement the display needs to be switched off manually, as we need to measure the reflection of the ambient light on the display panel.  You must place the sensor 15 cm away from the display facing the switched off panel.

After the acceptance test was performed, associated Constancy QA Tasks (tests) are automatically scheduled. You can now see them in the “Scheduler” tab.

PerfectLum QA Constancy Test (visual, performed bi-weekly) includes the following tests required by the NY PDM standard:

  1. SMPTE test pattern evaluation
  2. An all-white image evaluation 
  3. An all-black image evaluation
  4. Color trueness evaluation

PerfectLum QA Constancy Test (measurement, performed quarterly) includes the following test required by the NY PDM standard:

5. DICOM GSDF verification  

STEP 8. Perform Constancy tests when they are due

PerfectLum calculates the dates when constancy tests are due and will remind you to perform them, even if PerfectLum is not running on the date. So when you see PerfectLum Suite Task Notification window,  click the “Perform Scheduled Tasks” button and follow the on-screen instructions to evaluate the display performance.

STEP 9. Assign each test to the person responsible

According to the standard, each test must be assigned to an individual or a group of employees who will be responsible for it. For your convenience we have summarized the tests and their frequencies in the following table:

Name of the test Person/people Responsible Frequency How to perform
1

Acceptance Test including:

– Visual verification test

– White Uniformity test

– Black Uniformity test

– Color distortion test

– All displays White Uniformity test

– All displays Black Uniformity test

– DICOM GSDF verification test

– Luminance Response test

once,

when PerfectLum is installed

2

Constancy Test (visual) including:

– Visual verification test

– White Uniformity test

– Black Uniformity test

– Color distortion test

– All displays White Uniformity test

– All displays Black Uniformity test

Bi-Weekly

see step 8

3

Constancy Test including:

– DICOM GSDF verification test

– Luminance Response test

Quarterly

see step 8

STEP 10. Get ready for the annual evaluation

If you use PerfectLum and run Constancy tests (visual and measurement) on time, you will be ready for an annual evaluation by a Licensed Medical Physicist. The evaluation will include:

  1. DICOM GSDF verification
  2. Review of all facility QC documentation and procedures for PDMs.
  3. Quantitative assessment of the brightness (luminance) uniformity of each PDM
  4. Quantitative determination of the luminance ratio.
  5. Quantitative assessment of the viewing conditions.
  6. Analysis of all test results, including comparison with specifications provided by the vendor and applicable technical standards from professional societies
  7. Document recommendations for quality improvement and corrective actions.

The reports generated by PerfectLum after each operation can be submitted as QC documentation. In order to access the reports, open the  “History” tab and double-click on the necessary task. The detailed report will open as a .pdf file; you can print it or save it for future use.

ADDITIONAL INFORMATION

Acceptability limits of the  tests

  1. SMPTE test pattern evaluation (in PerfectLum: Visual Verification)
  1. An all-white image evaluation (in PerfectLum: White Uniformity and All displays White Uniformity)

On visual inspection, brightness should appear uniform on each PDM. If two or more PDMs are grouped to form a PDM workstation, there should be no noticeable difference in brightness between individual PDMs in the workstation.

  1. An all-black image evaluation (in PerfectLum: Black Uniformity and All displays Black Uniformity)

On visual inspection, the black should appear uniform on each PDM, and each PDM should have the same color tone. If two or more PDMs are grouped to form a PDM workstation, there should be no noticeable difference in brightness between individual PDMs in the workstation.

  1. Color trueness evaluation (in PerfectLum: Color distortion), only for PDMs capable of displaying color

The test pattern should contain objects of three colors, displaying shades of red, green, and blue. Objects should not be distorted.

Quarterly tests:

5. DICOM GSDF verification (in PerfectLum: Luminance Response and Luminance Response – DICOM curve).

For PDMs not used for mammography, the maximum luminance output of each PDM must be no less than 171 cd/m2, with a luminance ratio (defined as the ratio of maximum luminance to minimum luminance) no less than 170. Maximum luminance of at least 200 cd/m2 and a luminance ratio of 250 or greater are recommended.

For PDMs used for mammography, the maximum luminance output of each PDM must be no less than 250 cd/m2 , with a luminance ratio no less than 250. Maximum luminance of 450 cd/m2 and a luminance ratio of 500 or greater are recommended.

Equipment used for testing

You will only need a measurement device (puck). To check compatibility, please see step 2.

Sample records to be used for each test

DICOM Calibration for Improved Biomedical Imaging and Diagnostic Accuracy

In the fast-paced world of biomedical imaging, precision and accuracy are non-negotiable. From radiologists diagnosing critical conditions to technicians interpreting complex scans, every detail matters. That’s where DICOM Calibration steps in—not just as a technical requirement, but as the cornerstone of diagnostic confidence.

DICOM (Digital Imaging and Communications in Medicine) refers to the internationally recognized standard for medical imaging. It ensures that diagnostic displays reproduce grayscale images consistently, accurately, and in line with the specific luminance response defined in DICOM Part 14. For healthcare professionals, this is essential for interpreting everything from X-rays to MRIs without the risk of visual distortion or image misinterpretation.

Why DICOM Calibration Is Vital in Modern Radiology 

Medical images are not like regular photographs. A subtle difference in brightness, contrast, or grayscale could mean the difference between detecting a life-threatening abnormality—or missing it entirely. Without proper DICOM Calibration, even the most advanced and expensive display may not present images as intended by imaging equipment.

Key reasons calibration is critical include:

Introducing PL4 | The Next Generation of DICOM Calibration Software

If you’ve ever felt your premium medical display had untapped potential, you’re not alone. Many modern displays are capable of exceptional performance—but without proper software, much of that potential remains locked away.

That’s why QUBYX has developed PL4, our newly released DICOM Calibration suite designed to elevate your display to its highest capabilities.

What makes PL4 different?

PL4 in Action: Improving Biomedical Imaging Workflows

PL4 has been engineered to work flawlessly within high-demand radiology and telemedicine environments. Whether you’re operating in a large hospital network or a specialized imaging center, PL4 ensures that every diagnostic workstation delivers uniform, DICOM-compliant images.

Our internal testing and collaboration with radiology experts have resulted in significant improvements over previous calibration software, including:

Learn How to Maximize PL4: New Video Tutorial Series

We know that technology adoption is only successful when it’s easy to understand and use. That’s why we’ve launched a brand-new YouTube channel dedicated to helping you get the most from PL4.

Our short, easy-to-follow tutorials guide you step-by-step—from your very first calibration to advanced QA configurations. The first tutorial in the series shows exactly how to perform DICOM Calibration on your monitor using PL4 in just minutes.

Watch Now: [How to DICOM Calibrate Your Monitor with PL4]

Remote DICOM Calibration for the Teleradiology Era

With healthcare increasingly moving towards remote diagnostics, PL4 is equipped with a reliable remote calibration solution. This allows IT and QA teams to manage calibration centrally—ensuring every display, no matter the location, remains compliant with DICOM standards.

For remote radiologists, this means you can work from anywhere without compromising image quality or diagnostic reliability.

Try PL4 Today—It’s Free

Experience the power of next-level DICOM Calibration yourself. PL4 is available for free download, giving you access to professional-grade calibration tools designed to enhance biomedical imaging quality and diagnostic confidence.

[Download PL4 Now]
[Visit Our YouTube Channel]

Final Thoughts

At QUBYX, our mission is simple:
DICOM Calibration should be effortless. Compliance should be automatic. Image confidence should be guaranteed.

With PL4, we’ve combined advanced research, professional-grade performance, and user-friendly design into a single solution—helping radiology professionals unlock the full potential of their displays.

Read More:

Introduction to PerfectLum for Healthcare Professionals

PerfectLum4 is The Future of Medical Display Calibration Is Here

We are presenting the lastest PerfectLum4 software for medical disply calibration. In the field of medical imaging, precision is everything. From radiologists interpreting CT scans to technicians reviewing X-rays, every pixel matters. That’s why accurate DICOM calibration isn’t just a technical formality—it’s a clinical necessity.

At QUBYX, we understand this better than anyone. That’s why we’ve taken a significant leap forward with the release of PerfectLum4, our next-generation DICOM calibration software suite designed for professional-grade monitors and consumer-level displays alike.

Why DICOM Calibration Matters

Medical-grade image quality demands more than just good hardware. Without proper calibration, even the best monitors can deliver inaccurate results, leading to misinterpretations and compromised diagnoses. DICOM Part 14 compliance ensures that grayscale images are displayed consistently and correctly across all viewing environments.

PL4 makes this process simpler, smarter, and faster than ever before.

Meet PerfectLum4: The Most Advanced Version Yet

If you’ve ever felt like your premium monitor has untapped potential, you’re not wrong. The newly released PL4 calibration suite is specifically designed to unlock it. PerfectLum 4 is a professional-grade display calibration and quality‑assurance (QA) tool aimed primarily at medical imaging but flexible enough to support virtually any screen. Developed by QUBYX, it ensures that displays conform to critical standards such as DICOM GSDF, AAPM TG 18, DIN 6868‑57/157, JESRA X‑0093, IEC 62563‑1, and more.

It supports both Windows and macOS platforms and integrates with a wide array of measurement devices, graphics boards, and monitor models.

Key features include automated calibration, acceptance and constancy testing, scheduling with due‑date tracking, detailed reporting, and a remote QA server for centralized multi‑station management.

A notable milestone: PerfectLum 4 received FDA 510(k) clearance in 2020 when bundled with a 5K iMac as a diagnostic medical display solution.

With its ease of use, extensive compliance coverage, and robust remote capabilities, PerfectLum4 offers comprehensive tools for maintaining display accuracy in demanding clinical environments.

Key Advantages of PerfectLum4:

Easy to Use, Powerful in Performance

What makes PL4 stand out isn’t just the technology—it’s the experience. We’ve combined advanced research with real-world radiology workflows to create a tool that’s both powerful and accessible. Whether you’re calibrating a single diagnostic workstation or managing an entire fleet of displays, PerfectLum4 adapts to your needs.

Learning Made Simple: Video Tutorials Now Online

To help you get the most out of your PerfectLum 4 software, we’ve launched a dedicated YouTube channel. Our new tutorial series walks you through every essential step—from initial setup to advanced configuration and remote management.

Watch the First Tutorial Now:

Learn how to DICOM-calibrate your monitor with PL4 in just a few minutes.

Whether you’re new to calibration or upgrading from an earlier version, these bite-sized videos will get you up and running in no time.

Try PerfectLum4 Today—Absolutely Free

Ready to see the difference for yourself? The full version of PerfectLum 4 is available now for free. Join thousands of radiologists and IT administrators who are already discovering how effortless quality assurance can be.

[Download PL4]
[Visit Our YouTube Channel]

Conclusion

At QUBYX, we believe calibration should be effortless, compliance should be automatic, and image confidence should be guaranteed. With PerfectLum 4, we’re not just keeping up with the future of medical imaging—we’re helping shape it.

Try PerfectLum 4 today; it’s free.

Read More:

Introduction to PerfectLum for Healthcare Professionals

Benefits of Purchasing the PerfectLum4 DICOM Calibration Suite

In modern medical imaging, accuracy isn’t just an advantage—it’s a necessity. Radiologists and imaging professionals rely on diagnostic displays to reveal every fine detail, from the smallest fracture to the subtlest tissue abnormality. To achieve this level of precision, proper calibration is essential. That’s where PerfectLum4 DICOM comes in—an advanced, professional-grade calibration suite designed to unlock the full potential of your premium display.

Understanding DICOM Standards in Medical Imagine

DICOM (Digital Imaging and Communications in Medicine) defines the industrial standards for diagnostic displays, ensuring consistent, reliable grayscale reproduction across all viewing platforms. The DICOM Part 14 standard specifically governs the luminance response of medical-grade monitors, ensuring that each pixel is displayed with the correct brightness and contrast.

Without compliance to DICOM standards, even a high-end display can deliver inaccurate images—potentially impacting diagnostic accuracy. This is why PerfectLum4 DICOM is a game-changer for healthcare organizations, imaging centers, and teleradiology providers.

Why PerfectLum4 DICOM Is Different

Calibrating your monitor has never been easier or more effective. While many calibration solutions are available, PerfectLum4 DICOM stands apart thanks to:

Unlocking the Hidden Potential of Your Display

Have you ever purchased a premium monitor and thought, “This is good, but it could be better”? In many cases, you’re right—your hardware may have more performance to offer, but without proper calibration software, it remains untapped.

The PerfectLum4 DICOM suite was designed to bridge this gap. Using cutting-edge algorithms and in-depth research on display performance, we’ve engineered a solution that optimizes your monitor to deliver exceptional accuracy for medical imaging.

Key Benefits of Purchasing PerfectLum4 DICOM

1- Enhanced Diagnostic Confidence

Every grayscale value is perfectly aligned with DICOM standards, reducing diagnostic errors.

2- Regulatory Compliance Made Easy

Automated acceptance and constancy tests ensure displays are always audit-ready.

3- Improved Workflow Efficiency

Faster calibration cycles mean less downtime for busy imaging departments.

4- Remote QA for Multi-Site Operations

Centrally manage display performance across different locations without physical visits.

5- Longevity and Cost Savings

Regular calibration extends display lifespan, reducing the need for premature replacements.

Learning Made Simple: Our YouTube Tutorial Series

We understand that even the most advanced software is only as good as the user’s ability to operate it. To ensure our clients get the most out of PerfectLum4 DICOM, we’ve launched a dedicated YouTube channel.

Our new video tutorial series offers short, step-by-step guides covering everything from first-time setup to advanced remote QA workflows. The first tutorial focuses on performing a DICOM calibration with PerfectLum4—giving you hands-on knowledge in just minutes.

Watch Now: [How to Calibrate Your Monitor with PerfectLum4 DICOM]

Designed for the Future of Radiology

As medical imaging evolves, so do the demands placed on diagnostic displays. Technologies like high-resolution mammography, advanced CT imaging, and remote telehealth require consistent, compliant, and precise monitor calibration—every single time.

PerfectLum4 DICOM is built to meet these evolving challenges, ensuring that healthcare providers can deliver accurate diagnoses without worrying about display performance.

Try PerfectLum4 DICOM Today—Free

Experience the difference that PerfectLum4 DICOM can make for your workflow. Our full-feature trial is available free of charge, allowing you to explore its powerful features before making a purchase decision.

[Download PerfectLum4 DICOM]
[Watch Our Tutorials]

Final Thoughts

At QUBYX, our mission is clear:

Calibration should be effortless. Compliance should be automatic. Image confidence should be guaranteed.

With PerfectLum4 DICOM, we’ve created a tool that’s not only powerful but also practical—combining professional-grade accuracy with ease of use. Whether you’re a radiologist, a hospital IT administrator, or a teleradiology provider, PerfectLum4 DICOM will help you see your images as they were meant to be seen—accurately, consistently, and with complete confidence.

 
Read More:

Introduction to PerfectLum for Healthcare Professionals

PerfectLum4 Tutorial Series | Master DICOM Calibration with Ease

In the demanding world of medical imaging, there’s no room for guesswork when it comes to image quality. You musr read the PerfectLum4 Tutorial Series, whether you’re a radiologist, a technician, or an IT administrator. You know that the accuracy of diagnostic displays directly impacts patient care. That’s why calibration is more than just a technical requirement—it’s a clinical priority.

With the release of PerfectLum4, our latest DICOM calibration suite, we’ve taken professional display calibration to the next level. And to help users make the most of it, we’ve launched the PerfectLum4 Tutorial Series, an easy-to-follow video guide designed for every medical imaging professional.

Unlocking the Full Potential of Your Display

Have you ever bought a premium display and thought, “This is good, but I know it can do more”? You’re right—many modern monitors have untapped capabilities. Without proper calibration, even the most advanced displays can fall short of their potential.

The PerfectLum4 calibration suite is designed to unlock that hidden performance. Built on advanced algorithms and extensive research into the latest display technologies, PerfectLum4 ensures your screen meets DICOM Part 14 standards—a critical requirement for accurate grayscale rendering in medical imaging.

Why the PerfectLum4 Tutorial Series Matters

Even though PerfectLum4 is designed to be intuitive, we understand that medical professionals need quick, precise guidance. That’s why we created the PerfectLum4 Tutorial Series—a collection of short, practical video lessons available on our official QUBYX YouTube channel.

This series ensures that whether you’re new to calibration or a seasoned IT administrator, you can:

What’s in the PerfectLum4 Tutorial Series?

The tutorials are structured to give you exactly what you need, when you need it, including:

1- Introduction to PerfectLum4

Overview of features, benefits, and how it integrates into medical imaging workflows.

2- Step-by-Step DICOM Calibration

How to measure, adjust, and verify grayscale luminance.

Ensuring your display meets DICOM Part 14 compliance.

3- Display Conformance Verification

Running automated acceptance and constancy tests.

Creating and exporting compliance reports for regulatory audits.

4- Remote QA Management

Setting up and maintaining display calibration across multiple locations.

Centralized monitoring for large-scale healthcare networks.

5- Advanced Customization

Tailoring calibration schedules and QA protocols to your facility’s needs.

PerfectLum4: Built for Today’s and Tomorrow’s Radiology Needs

As telemedicine and teleradiology continue to expand, the need for accurate, remotely managed display calibration is greater than ever. PerfectLum4’s remote QA capabilities allow calibration teams to ensure every monitor—whether in a hospital reading room or a remote radiologist’s home office—meets the same high standards.

With full Windows 11 compatibility, PerfectLum4 integrates seamlessly into modern IT infrastructures, providing stability, speed, and reliability.

Benefits of Using PerfectLum4 with the Tutorial Series

When you combine the power of PerfectLum4 with the knowledge from the tutorial series, you get:

How to Access the PerfectLum4 Tutorial Series

Getting started is simple:

1- Download PerfectLum4 – Try the software for free and explore its full range of features.

2- Visit Our YouTube Channel – Watch the PerfectLum4 Tutorial Series playlist and follow along step-by-step.

3- Start Calibrating – Apply what you learn immediately and experience the improvement in image accuracy.

First Video Available Now: How to DICOM Calibrate Your Monitor with PerfectLum4

Try PerfectLum4 Today—It’s Free

You don’t have to wait to experience the difference. Download the free trial of PerfectLum4 today and start using the PerfectLum4 Tutorial Series to achieve professional-grade calibration with ease.

[Download PerfectLum4]
[Watch the Tutorial Series]

Final Thoughts

At QUBYX, we believe that calibration should be effortless, compliance should be automatic, and diagnostic confidence should be guaranteed. With the PerfectLum4 Tutorial Series, we’ve made it easier than ever for medical professionals to achieve and maintain the highest standards in display performance.

Whether you’re managing a single workstation or an entire network of displays, PerfectLum4—backed by our tutorial series—puts precision calibration in your hands.

 